Relationship between radiation dose and reduced X-ray sensitivity surrounding breast region using CR stimulable phosphor plate for mammography

机译:辐射剂量与乳房区乳房X线摄影辐射磷光区辐射剂量和X射线敏感性的关系

获取原文

摘要

Computed radiography (CR) systems use a photostimulable phosphor plate (imaging plate ; IP) as a sensor for digital mammography. In clinical mammography, breast is almost exposed same region of IP, and therefor, direct x-ray regions surrounding suffer from reduced x-ray sensitivity. Consequently, the difference in x-ray sensitivity between the breast regions and the unattenuated x-ray region was obtained. However, radiation dose quantity that reduces x-ray sensitivity is not known. In this study, we imaged a breast phantom under fixed conditions, and subsequently, we investigated the pixel value differences between the breast region and the unattenuated x-ray regions. We measured the entrance air-kerma using 550 sensing elements of glass dosimeter, 22x25 lines, that were placed at the surface of the cassette including the IP. In order to measure the x-ray sensitivity, pre- and post-exposure breast phantom images were acquired after 500, 1,000, 1,350, and 1,500 trials. The pixel values were measured at four points; in the breast region and in the unattenuated x-ray region. The ratio of these pixel values was compared with the cumulative exposure dose. The ratio was nearly constant until 1,000 trials, but a significant reduction was observed after 1,350 trials. Further, in the image obtained after 1,500th trials, the shape of breast phantom could be observed. This image supports the fact that the x-ray sensitivity was lowered in the unattenuated x-ray region. The difference in the pixel value between the breast region and the unattenuated x-ray region was obtained over 1,000 exposures at 100,000 mAs.
机译:计算的射线照相(CR)系统使用光刺激的磷光板(成像板; IP)作为数字乳房X光检查的传感器。在临床乳房爆炸中,乳房几乎暴露相同的IP区域,而且,周围的直接X射线区域患有降低的X射线敏感性。因此,获得了乳房区和未延长的X射线区域之间的X射线敏感性的差异。然而,不知道降低X射线敏感性的辐射剂量量是未知的。在这项研究中,我们在固定条件下成像乳房幻影,随后,我们研究了乳房区域和未滞后的X射线区域之间的像素值差异。我们使用玻璃剂量计的550个感测元件测量进入空气 - KERMA,玻璃剂量计,22x25线,置于包括IP的盒的表面。为了测量X射线敏感性,在500,1,000,1,350和1,500次试验后获得预曝光前和暴露后乳房幻影图像。在四个点测量像素值;在乳房区域和未延强的X射线区域中。将这些像素值与累积曝光剂量进行比较。该比率几乎是恒定的,直至1,000试验,但在1,350次试验后观察到显着的减少。此外,在第1500次试验后获得的图像中,可以观察到乳房幻影的形状。该图像支持在未滞后的X射线区域中降低了X射线灵敏度的事实。在100,000mas下获得超过1,000个曝光的乳房区域和未滞后X射线区域之间的像素值的差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号