首页> 外文会议>Biomedical imaging and sensing conference >Study of noncontact air-puff applanation tonometry IOP measurement on irregularly-shaped corneas
【24h】

Study of noncontact air-puff applanation tonometry IOP measurement on irregularly-shaped corneas

机译:非接触式空气 - 浮粉型肌室测量对不规则形状的角膜的研究

获取原文

摘要

Abnormal corneas with corneal tissue defects like ulceration, melting, laceration, thinning scar, keratoconus etc., poses special challenges for ophthalmologist to measure intraocular pressure (IOP) correctly using air-puff noncontact applanation tonometry. Here, we propose an novel model, Abnormal Applanation IOP Model (AAIOP), to simulate IOP in these abnormal corneas on an air-puff noncontact applanation tonometry system, and the simulated IOP results are correctly fit in those of IOP measured database on human eyes of 91,024 patients (174,666 eyes). Our simulated IOP indicates that every 10 μm of central corneal thickness change results in 0.36 mmHg of IOP change. Using our simulation model, the IOP on abnormal eyes with irregularly-shaped corneas can be correctly expected and reported.
机译:角膜异常具有角膜组织缺陷,如溃疡,熔化,撕裂,稀释疤痕,角蛋白酶等,使用空气 - 浮气非接触式施加质量测量眼科医生的特殊挑战来测量眼压(IOP)。在这里,我们提出了一种新颖的模型,异常的施防IOP模型(AAIOP),在空气 - 浮烟非接触式施加型系统上模拟这些异常角膜中的IOP,并且模拟的IOP结果在人眼中的IOP测量数据库中正确适合91,024名患者(174,666只眼)。我们的模拟IOP表示每10μm的中央角膜厚度变化导致0.36mmHg的IOP变化。使用我们的仿真模型,可以正确地预期并报告具有不规则形状的角膜异常眼睛的IOP。