摘要

In this work, a new photoplethysmography (PPG) signal processing methodology using polynomial profile fitting for measuring blood flow volume (BFV) in arteriovenous fistula (AVF) is proposed. After calibration, the experiment results using proposed method shows higher correlation (R=0.7883) and lower error (RMSE=109 ml/min) compared to the ones using conventional methods (R=0.3212, RMSE=168 ml/min).
机译:在这项工作中,提出了一种新的利用多项式轮廓拟合的光体积描记术(PPG)信号处理方法,用于测量动静脉瘘(AVF)中的血流量(BFV)。校准后,与使用常规方法(R = 0.3212,RMSE = 168 ml / min)相比,使用建议方法的实验结果显示出更高的相关性(R = 0.7883)和更低的误差(RMSE = 109 ml / min)。

著录项

  • 来源
    《》|2017年|V001T09A003.1-V001T09A003.3|共3页
  • 会议地点 San Francisco(US)
  • 作者

    Paul C.-P. Chao; Pei-Yu Chiang;

  • 作者单位

    Department of Electrical Engineering National Chiao Tung University, Hsinchu, Taiwan;

    Department of Electrical Engineering National Chiao Tung University, Hsinchu, Taiwan;

  • 会议组织