This paper presents the results of safety and performance inspection conducted on 325 infusion pumps in the period from January 2015 to January 2016 in public and private healthcare institutions in Bosnia and Herzegovina. Electrical safety inspection was conducted according to IEC 60601-Medical electrical equipment: General requirements for basic safety and essential performance of medical devices. Periodical performance inspections were conducted according to legal metrology framework, more specifically according to Rules on metrological and technical requirements for infusion pumps, published in Official Gazette of Bosnia and Herzegovina (BH) No. 75/14. Our results show that approximately 9% of all tested infusion pumps do not meet electrical safety requirements or have performance outside the specifications. The causes of such faulty performance are environmental conditions, long period of usage of devices, faulty motors and device casing, as well as lack of systematic maintenance and regular safety and performance inspections in previous period. This paper addresses the importance of periodical performance inspections of these medical devices.
机译:本文介绍了2015年1月至2016年1月在波斯尼亚和黑塞哥维那的公共和私立医疗机构的325个输液泵的安全性和绩效检查结果。电气安全检验根据IEC 60601-医用电气设备进行:一般要求基本安全和医疗设备的必要性能。根据法律计量框架进行的定期绩效检查,更具体地说,根据灌注泵的计量和技术要求的规则,在波斯尼亚和黑塞哥维那(BH)第75/14号官方公报中发表。我们的研究结果表明,大约9%的所有测试的输液泵都不符合电气安全要求或在规格之外的性能。这种故障性能的原因是环境条件,长期使用设备,故障电机和器件套管,以及前期缺乏系统维护和定期安全性和性能检查。本文涉及这些医疗设备的期刊性能检查的重要性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号