首页> 外文会议>International Workshop on ADC and DAC Modelling and Testing;IMEKO TC4 Symposium Measurement of Electrical Quantities;Symposium IMEKO TC19 - Environmental Measurements >A Methodology to Assess Immunity to Radiated Disturbances of Fpaa-Based Front-End Devices for Electrocardiograph Signal Monitoring
【24h】

A Methodology to Assess Immunity to Radiated Disturbances of Fpaa-Based Front-End Devices for Electrocardiograph Signal Monitoring

机译:一种评估来自基于FPAA的前端装置的辐射干扰的方法,用于心电图信号监控

获取原文

摘要

The paper is focused on the assessment of the functional immunity of field programmable analog arrays (FPAA) circuit configured to operate as front-end elements to be used in electrocardiography (ECG) signal conditioning, and on the investigation of the possible effects of radiated electromagnetic disturbance in the range from 250 MHz to 3 GHz, on the parameters of the ECG signal. The research is carried out with a methodology which relies on typical ECG signal and both time and frequency domain analysis. Results show that under specific amplitude and frequency conditions, the disturbance can indeed interfere with the normal operation, leading to erroneous diagnosis.
机译:本文重点研究了现场可编程模拟阵列(FPAA)电路的功能性免疫的评估,该电路被配置为作为用于心电图(ECG)信号调节的前端元件,以及对辐射电磁的可能影响的研究 在ECG信号的参数上,从250 MHz到3 GHz的扰动范围内。 该研究具有依赖于典型的ECG信号和时间和频域分析的方法进行的方法进行。 结果表明,在特定幅度和频率条件下,干扰确实会干扰正常操作,导致错误诊断。