首页> 外文会议>Human Factors and Ergonomics Society international annual meeting >The Usability of Blood Glucose Meters: Task Performance Differences Between Younger and Older Age Groups
【24h】

The Usability of Blood Glucose Meters: Task Performance Differences Between Younger and Older Age Groups

机译:血糖计的可用性:年轻人和旧年龄组之间的任务性能差异

获取原文

摘要

This study examines the overall usability of current, commercially available self-monitoring blood glucose (SMBG) meters in order to highlight how the design affects the performance of younger and older adults. Sixteen younger participants (18-27) and 29 elderly participants (65-85) attempted to complete two tasks: 1) set date/time and 2) perform a control solution test, using two meters: the Accu-Check Compact Plus? and 2) the One Touch Ultra 2?. Overall, elderly participants had significantly more difficulty completing the two tasks and committed significantly more errors than younger participants. When using the Accu- Chek? meter to set the date and time 79% of elderly adults failed compared to 12% of younger adults. When using the One Touch? meter, 21% of elderly adults failed to complete the task while none of the younger adults failed. Across the two tasks, elderly participants made, on average, an additional 1.6 errors compared to those in the younger group.
机译:本研究探讨了当前,商业上可用的自我监测血糖(SMBG)米的总体可用性,以突出设计如何影响年轻人和老年人的表现。十六名年轻的参与者(18-27)和29名老年参与者(65-85)试图完成两个任务:1)设定日期/时间和2)执行控制解决方案测试,使用两米:ACCU-CHECK COMPACT PLUS? 2)一个触摸超2?总体而言,老年人参与者难以完成两项任务的难度越来越突出,而且犯下比年轻参与者更大的错误。使用Accu-Chek时?仪表设定日期和时间79%的老年人的成年人失败,而占年幼的成年人的12%。使用单一触摸?仪表,21%的老年人未能完成任务,而年轻人则没有失败。在两位任务中,年长参与者平均额外的1.6次错误与年轻群体相比。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号