Dynamic Electrocardiograph (ECG) monitoring (known as Holter) plays an important role in the earlier detection and diagnosis of various cardiovascular diseases. ECG signals obtained from Holter systems normally contain a lot of noises and artifacts. These noises degrade signal quality, which may be critical for routine monitoring and diagnosis. To solve the problem, a reconfigurable overlapping fast Fourier transform/ inverse fast Fourier transform (FFT/IFFT) filter for suppressing the power-line interference and the high-frequency noise is presented in this paper. The filter is based on a 12-lead Holter system with a high-performance analogue front-end and a field-programmable gate array (FPGA) for enhanced digital processing. This paper analyzes the performance of the reconfigurable overlapping FFT/IFFT filter in ECG de-noising applications and validate it by real-world emulations. Furthermore, the de-noising performance of the reconfigurable overlapping FFT filter was evaluated.
机译:动态心电图(ECG)监测(称为HOLTER)在早期的检测和诊断中发挥着重要作用,诊断各种心血管疾病。从HOLTER系统获得的ECG信号通常包含大量噪声和伪影。这些噪声降低了信号质量,这可能对常规监测和诊断至关重要。为了解决问题,本文提出了一种用于抑制电力线干扰的可重构重叠的快速傅里叶变换/逆快速傅里叶变换(FFT / IFFT)滤波器和高频噪声。该滤波器基于12引导孔系统,具有高性能模拟前端和用于增强数字处理的现场可编程门阵列(FPGA)。本文分析了ECG去噪应用中可重新配置重叠FFT / IFFT过滤器的性能,并通过现实世界仿真验证。此外,评估可重构重叠的FFT滤波器的去噪性能。