首页> 外文会议>International Seminar on Biomaterials and Regenerative Medicine >Titanium Implants used in Traumatology Failed Sometime in Clinical Practice: Case Study
【24h】

Titanium Implants used in Traumatology Failed Sometime in Clinical Practice: Case Study

机译:在临床实践中有时使用的钛植入物失败:案例研究

获取原文

摘要

Tibial fractures were, still are and will be a challenge for the orthopedic surgeon. In this paper we conducted a case study on a system plate-screw made by titanium that was used in clinical practice for reduction and fixation of a tibial pilon fracture. After eight months the patient returns for the ablation of metallic implants. Clinical and strengthen the fracture callus is found radiologic hypertrophic and degradation plaque.On retrieved plate were made following investigations in order to establish the causes that lead to the failure: determining the chemical composition through spectral analysis, macrostructural analysis using stereomicroscopy, microstructure analysis using optical microscope metallographic, and fractographic analysis using scanning electron microscopy. Complex analysis of the fracture surfaces of the locking plate has led to the ultimate conclusion that the material has been made self-locking plate is inadequate chemical purity satisfactory embedding, which has led to breakage within an area of non-homogeneous structure. Breaking looks fragile behaviour by transgranular cleavage with sharp facets. There were also highlighted intergranular cracks between planes of cleavage sides.
机译:胫骨骨折是,仍然是骨科外科医生的挑战。在本文中,我们对钛制成的系统板螺杆进行了案例研究,用于减少和固定胫骨皮骨折的临床实践。八个月后,患者返回金属植入物的消融。临床和加强骨折愈伤组织被发现放射生理肥厚和降解斑块。在调查中进行了检索的板,以便建立导致失败的原因:通过光谱分析确定化学成分,使用立体镜检查,使用光学分析进行微观结构分析。使用光学分析显微镜金相和使用扫描电子显微镜的式分析。对锁定板的断裂表面的复杂分析导致了最终的结论,即该材料已成为自锁板是不充分的化学纯度令人满意的嵌入,这导致了非均匀结构区域内的破损。通过用尖锐的刻面进行响囊裂解,破坏了脆弱的行为。在裂解侧面之间也存在突出的晶间裂缝。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号