首页> 外文会议>International Conference on Frontiers of Mechanical Engineering and Materials Engineering >The Analysis of Bioprosthetic Heart Valve with Different Suture Densities
【24h】

The Analysis of Bioprosthetic Heart Valve with Different Suture Densities

机译:不同缝合密度的生物假体心瓣分析

获取原文
获取外文期刊封面目录资料

摘要

In order to investigate the effect of suture density on the dynamic behavior of bioprosthetic heart valve, we establish the spherical leaflets models via computer aided design. Based on the parametric models of the heart valve, four kinds of suture density is analyzed and finite element analysis is used to simulate the mechanical performance of bioprosthetic heart valve. The results show that the stress distributions of the spherical valve leaflets with different suture density is quite different and the peak von-Mises with 50 suture points is lower than the other three kinds of suture density. From the whole loading process, we can find that the spherical valve leaflet with 50 suture points has better dynamic properties. This work is very helpful when manufacuturing the bioprosthetic heart valve, thus to prolong the lifetime of the bioprosthetic heart valve.
机译:为了探讨缝合密度对生物假物瓣膜动态行为的影响,我们通过计算机辅助设计建立了球形小叶模型。基于心脏瓣膜的参数模型,分析了四种缝合密度,有限元分析用于模拟生物假体心脏瓣膜的机械性能。结果表明,具有不同缝合密度的球形瓣膜小叶的应力分布是完全不同的,并且具有50个缝合点的峰值峰值低于其他三种缝合密度。从整个装载过程中,我们可以发现具有50个缝合点的球形阀门宣传片具有更好的动态性能。这项工作在制造生物假体心脏瓣膜时非常有帮助,从而延长生物假体心脏瓣膜的寿命。