A fine tube type glucose sensor, which has a sensing region at the inside wall of fine tube tip, was prepared and applied for minimally invasive glucose monitoring. The enzyme glucose oxidase was immobilized on the platinum surface located in the tube by the combination of electrodeposition and electro-polymerization of o-phenylenediamine. Amperometric responses of the prepared electrodes to glucose were measured at a potential of 0.60 V (vs Ag/AgCl). The tube type electrode showed good response with good linear relationship up to the glucose concentration of 22.4 mM, which was actually higher than normal sensor without outer film for permeability restriction. This may be due to its unique sensor structure.
机译:制备具有在细管尖端的内壁的细管式葡萄糖传感器,并施加用于微创葡萄糖监测。通过电沉积和电聚合的O-苯二胺的组合,将酶葡萄糖氧化酶固定在位于管中的铂表面。在0.60V(Vs Ag / AgCl)的电位下测量制备的电极对葡萄糖的安培响应。管型电极显示出良好的响应响应良好的线性关系,直到22.4mm的葡萄糖浓度,其实际上高于没有外膜的正常传感器,用于渗透限制。这可能是由于其独特的传感器结构。