Timing Stability of TCSPC Experiments

机译:TCSPC实验的时序稳定性

获取原文

摘要

Advanced time-correlated single-photon counting (TCSPC) devices are able to record several 106 photons per second and deliver an instrument response function down to 25 ps FWHM. Under these conditions the accuracy of fluorescence decay or photon migration times is limited by systematic timing errors rather than by the photon statistics. The experiments described below determined the variation of the instrument response function (IRF) with the count rate and the timing drift for an SPC-140 TCSPC module and a number of commonly used detectors. For count rates from 3·10~4 to 4·10~6 s-1 a shift of the first moment of the IRF smaller than 2 ps was obtained. The drift over 16 minutes was within ±0.7 ps.
机译:高级时间相关的单光子计数(TCSPC)设备能够每秒记录多个106光子,并将仪器响应函数降至25 ps fwhm。在这些条件下,荧光衰减或光子迁移时间的准确性受系统定时误差而不是光子统计的限制。下面描述的实验确定了仪器响应函数(IRF)与SPC-140 TCSPC模块的计数率和定时漂移的变化以及许多常用的检测器。对于3·10〜4至4·10〜6 S-1的计数率,获得了小于2 PS的IRF的第一矩的偏移。 16分钟超过16分钟的漂移在±0.7 ps内。