Novel single x-ray absorptiometry method to solve for volumetric breast density in mammograms with paddle tilt

机译:用桨倾斜解决乳房X线图中的体积乳房密度的新型单X射线吸收法

获取原文

摘要

We report on the design and validation of a breast tissue equivalent phantom for automated measurement of breast composition in film/screen and digital mammography systems. This phantom is a multi-step phantom made of a single material and containing nine lead positioning markers around its periphery. The markers allow for the phantom position to be solved relative to the x-ray gantry. The phantom was adhered to the top of the mammographic compression paddle such that it projected an attenuation image onto the unused corner of the image without overlapping with CC- nor MLO-view breast edges. The markers and their centroids were identified using automatic morphological image processing operations. The phantom, and thus the paddle, orientation is then obtained by minimizing a simple least-square error function of the difference between a pseudo projection image of the phantom markers at known coordinates and the actual marker image. Fibroglandular-equivalent breast attenuation values were found directly from step phantom projections. Fat attenuation values were derived from the attenuation coefficient ratios of fat to fibroglandular tissue. Finally, breast density was calculated by comparing image pixel values to the fat/fibroglandular references at the same thickness. Multiple scans of a test object (a density step phantom with 7 densities) at nine different compression thicknesses and six paddle-tilt angles were acquired. We found the precision for determining the breast thickness to be 0.015 cm (standard deviation) and for determining individual paddle angles to be 0.05 degrees. Multiple clinical studies using the technique on film/screen and digital mammography machines are also currently under way.
机译:我们报告了薄膜/屏幕和数字乳房X线摄影系统中乳房组合物自动测量的乳房组织等同模体的设计和验证。该幻影是由单个材料制成的多步体,并围绕其周边含有九个铅定位标记。标记允许相对于X射线龙门架求解幻象位置。幻影被粘附到乳房X线切压缩桨的顶部,使得它将衰减图像投影到图像的未使用的角落上而不与CC-也不重叠,也不重叠。使用自动形态图像处理操作识别标记及其质心。然后,通过最小化已知坐标和实际标记图像的幻像标记的伪投影图像之间的差异的差异的简单最小二乘误差函数来获得幻像。直接从步进幻影投影中发现纤维族相当的乳房衰减值。脂肪衰减值源自脂肪脂肪族对纤维绦虫组织的衰减系数比。最后,通过将图像像素值与以相同厚度的脂肪/纤维族引用进行比较来计算乳房密度。获取在九个不同压缩厚度和六个桨倾斜角度下的测试对象(具有7密度的密度步体幻象)的多次扫描。我们发现确定乳房厚度为0.015cm(标准偏差)的精度,并且用于将单个桨角确定为0.05度。使用该技术采用薄膜/屏幕和数字乳房X线摄影机器的多种临床研究也在正在进行中。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号