With the purpose of generating the curiosity of the public, a pulse-tube cryocooler with regenerator, pulse-tube, inertance tube and reservoir made of glass has been designed constructed and operated. The dimensions of the glass regenerator have been determined using REGEN3.3 [1] from given parameters of the conductive porous medium inside of the regenerator and a 150K target cooling temperature at the cold head. The geometry of the glass pulse-tube and glass inertance tube has been fixed using an approximate design method [2], and the entire system parameters checked using SAGE [3]. The thickness of each glass component is based on a charge pressure of around 7 bar and apressure ratio of about 1.35. The dimensions of the after-cooler are calculated using ISOHX [4] assuming a gas temperature of 300 K at the inlet of the regenerator.