To quantitatively evaluate the performance of a non-linear chirp compression technique, we made b-mode images of a flow phantom using pulse and chirp excitations in both fundamental and second harmonic mode. Flow channel sizes as measured from the obtained images show good agreement between the imaging methods, with only a small loss in axial resolution for chirp harmonic imaging. However, range side-lobes were present in the chirp harmonic images. SNR improvements of up to 11 dB were found for chirp excitation relative to pulse excitation.
机译:为了定量评估非线性啁啾压缩技术的性能,我们在基本和二次谐波模式下使用脉冲和啁啾激发来制造流模型的B模式图像。从所获得的图像测量的流动通道尺寸显示成像方法之间的良好一致性,仅在啁啾谐波成像的轴向分辨率下损耗很小。然而,啁啾谐波图像中存在范围侧瓣。相对于脉冲激发,发现啁啾激励最多11dB的SNR改善。