首页> 外文会议>Joint Engineering in Medicine and Biology Society-Biomedical Engineering Society Conference >The assessment of therapeutic ultrasound heating by microwave radio-thermometer measurements
【24h】

The assessment of therapeutic ultrasound heating by microwave radio-thermometer measurements

机译:微波无线计测量评估治疗超声热量测量

获取原文

摘要

A brightness temperature dissipated by biological tissue in vivo and in vitro was measured using a single-frequency Microwave Radio-Thermometer (MRT). The temperature measurement in vivo and in vitro was performed right after Therapeutic Ultrasound Heating (TUH) energy was applied to a specimen of pork and a canine thigh, respectively. Throughout the statistical regression analysis between the brightness temperature measured by MRT and internal one sensed by a thermocouple sensor, we concluded that MRT could well resolve the internal temperature of biological tissues in terms of an average temperature within a volume-of-interest coupled with MRT antenna.
机译:使用单频微波无线电温度计(MRT)测量通过体内和体外体外散发的亮度温度。在治疗超声加热(Tuh)能量分别施加到猪肉和犬大腿标本后,在体内和体外体内测量。在通过热电偶传感器测量的亮度温度之间的阳性回归分析中,我们得出结论,MRT可以在与MRT的兴趣体积内的平均气温方面重新解决生物组织的内部温度天线。