In cone-beam 3D CT, 2D X-ray projections determine the reconstruction quality directly. However, the generation of an X-ray projection is a complex process that includes the emission, the conversion and the transfer of X-ray latent information, so the pixel value of a projection has no exact meaning generally. In this paper, we suggest a way by experiments to model the whole imaging system based on "equivalent single energy" which is used to analyze polychromatic X-ray. We define each tissue's SCF(single calibration factor) of human and get GCF(gross calibration factor) by summing up every weighted SCF, then we can use GCF to calibrate and linearize X-ray projections. After such processing, X-ray projection will have explicit meaning, i.e. its pixel value can denote the equivalent material thickness linearly, which will be advantageous to the algorithm research of 3D reconstruction and other quantitative analyses of human's physiological parameters.
机译:在锥形光束3D CT中,2D X射线投影直接确定重建质量。然而,X射线投影的产生是一种复杂的过程,其包括发射,转换和X射线潜在信息的传送,因此投影的像素值通常没有精确的含义。在本文中,我们建议通过实验基于“等效单能”来模拟整个成像系统的方法,其用于分析多色X射线。我们通过每次加权SCF总结,定义人类的每个组织的SCF(单校准因子)并获得GCF(总校准因子),然后我们可以使用GCF来校准和线性化X射线投影。在这样的处理之后,X射线投影将具有明确的含义,即其像素值可以线性地表示相同的材料厚度,这将是对3D重建和人类生理参数的其他定量分析的算法研究。