首页> 外文会议>IEEE Nuclear Science Symposium Conference >CdZnTe Detectors for the Positron Emission Tomographic Imaging of Small Animals
【24h】

CdZnTe Detectors for the Positron Emission Tomographic Imaging of Small Animals

机译:Cdznte探测器用于小动物的正电子发射断层图像

获取原文

摘要

We report the timing performances between two cadmium zinc telluride (CZT) detectors equipped with a specific electrode geometry and a dedicated electronic setup for Positron Emission Tomography (PET) application. The measured coincidence times reach 2,6ns Full Width at Half Maximum (FWHM) for a 500V Mas voltage and 300keV energy threshold. Subsequently, a simulation study was carried out to assess the spatial and efficiency performances of these detectors which allow the Depth Of Interaction (DOI) measurement. Preliminary results show that the proposed design could reach a resolution better (1mm Pull Width at Half Maximum instead of 2,5mm in the middle of the field of view) than what is achieved with the standard PET scintillation detectors. Experimental confirmation of this result is in progress.
机译:我们报告了配备有特定电极几何形状的碲化镉(CZT)检测器之间的时序性能,以及用于正电子发射断层扫描(PET)应用的专用电子设置。测量的巧合时间为500V MAS电压和300keV能量阈值达到26ns的半最大(fwhm)的全宽度。随后,进行了模拟研究以评估这些探测器的空间和效率性能,其允许相互作用的深度(DOI)测量。初步结果表明,所提出的设计可以达到更好的分辨率(在视野中的半最大而不是2,5mm的1mm拉宽,而不是2,5mm的视野),而不是用标准的PET闪烁探测器实现的。实验证实这一结果正在进行中。