首页> 外文会议>Conference on atmospheric propagation and remote sensing >Remote gas detection and quantitative analysis from infrared emission spectra obtained by Fourier transform infrared spectroscopy
【24h】

Remote gas detection and quantitative analysis from infrared emission spectra obtained by Fourier transform infrared spectroscopy

机译:通过傅里叶变换红外光谱获得的红外发射光谱远程气体检测和定量分析

获取原文

摘要

Techniques for obtaining quantitative values of the temperatures and concentrations of remote hot gaseous effluents from their measured passive emission spectra have been examined in laboratory experiments. The high sensitivity of the spectrometer in the vicinity of the 2397 cm$+$MIN@1$/ band head region of CO$- 2$/ has allowed the gas temperature to be calculated from the relative intensity of the observed rotational lines. The spatial distribution of the CO$-2$/ in a methane flame has been reconstructed tomographically using a matrix inversion technique. The spectrometer has been calibrated against a black body source at different temperatures and a self absorption correction has been applied to the data avoiding the need to measure the transmission directly. Reconstruction artifacts have been reduced by applying a smoothing routine to the inversion matrix.
机译:实验室实验中,研究了从其测量的无源发射光谱获得温度和远程热气态流出物的温度和浓度的定量值的技术。 Co $ - 2 $ /最高@ 1 $ /带头区域附近的光谱仪的高灵敏度允许从观察到的旋转线的相对强度计算气体温度。使用矩阵反转技术在甲烷火焰中重建了CO $ -2 $ /在甲烷火焰中的空间分布。光谱仪在不同温度下针对黑色体源校准,并且已经应用​​于避免直接测量变速器的数据来校准自吸收校正。通过将平滑例程应用于反演矩阵来减少重建伪影。