Measurement of peroxyacetyl nitrate NO2 and NOx by using a gas chromatograph with a luminol-based detector

机译:使用气相色谱仪用鲁米诺的探测器测量过氧乙酰硝酸酯NO2和NOx的测量

获取原文

摘要

A commercial instrument for measuring PAN (peroxyacetyl nitrate), NO$- 2$/, and optionally, NOx, is described. It is based on a gas chromatograph (GC), and uses a Luminox$+R$/ luminol detector following conversion of the PAN to NO$-2$/. A detection limit (based on a peak height of 3 $MUL noise) of 30 pptV is achieved for injections made every five minutes. The instrument is computer driven and contains its own peak integrator and graphics output. No interferences have been identified, including chlorinated hydrocarbons, NO$-2$/ and ozone. However, the PAN analog peroxypropanyl nitrate (PPN) is not fully resolved from PAN. The maintenance interval is monthly. The calibration of the instruments is relatively easy: known mixing ratios of NO$-2$/ in air are used to calibrate the sample injection system and Luminox$+R$/ detector. The passage of PAN through the column and the subsequent conversion of PAN to NO$-2$/ has been found to be stable over times of one year, and is reproducible between instruments. Atmospheric measurements for PAN have been carried out for both very polluted environments and very clean environments. The results from a month long formal intercomparison between the Luminox$+R$/ PAN and two GCs equipped with electron capture detectors is presented. Those results demonstrate that the luminol based PAN detector can yield accurate measurements of ambient PAN concentrations. A method for calibrating PAN instruments in the field is described.
机译:描述了一种用于测量锅(过氧乙酰硝酸酯),NO $ - 2 $ /,可选的商业仪器,并且是NOx。它基于气相色谱仪(GC),并在将PAN转换为NO $-2 $ /。在每五分钟内进行的注射实现检测极限(基于3 $ MUL噪声的峰值高度)。仪器是计算机驱动的,并包含自己的峰积分器和图形输出。没有发现干扰,包括氯化烃,没有-2美元/和臭氧。然而,PAN模拟过氧丙二烷基硝酸盐(PPN)没有完全从锅中分离。维护间隔是每月的。仪器的校准相对容易:已知的MOTER-2 $ / IN空气中的混合比用于校准样品喷射系统和Luminox $ + R $ /探测器。平底锅通过柱子的通过和随后转换为没有$ -2 $ /已被发现在一年内稳定,并且在仪器之间可重复。对于非常污染的环境和非常干净的环境,已经为PAN进行了大气测量。提出了一个月的Luminox $ + r $ / pan与配备电子捕获探测器的两个GCS之间的一个月长的正式相互作用的结果。这些结果表明,基于鲁米诺的PAN检测器可以屈服于环境平移浓度的准确测量。描述了一种用于校准场中的校准仪器仪器的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号