The purpose of this study was to determine the Contrast to Noise Ratio (CNR) of the x-ray images taken with the phase contrast imaging mode and compare them with the CNR of the images taken under the conventional mode. For each mode, three images were taken under three exposure conditions of 100 kVp (2.8mAs), 120 kVp (1.9mAs) and 140kVp (1.42mAs). A 1.61cm thick contrast detail phantom was used as an imaging object. For phase contrast, the source to image detector distance (SID) was 182.88 cm and the source to object (SOD) distance was 73.15 cm. The SOD was the same as SID in the conventional imaging mode. A computed radiography (CR) plate was used as a detector and the output CR images were converted to linear form in relation with the incident x-ray exposure. To calculate CNR, an image processing software was used to determine the mean pixel value and the standard deviation of the pixels in the region of interest (ROI) and in the nearby background around ROI. At any given exposure condition investigated in this study, the CNR values for the phase contrast images were better as compared to the corresponding conventional mode images. The superior image quality in terms of CNR is contributed by the phase-shifts resulted contrast, as well as the reduced scatters due to the air gap between the object and the detector.
机译:本研究的目的是确定与相位对比度成像模式拍摄的X射线图像的噪声比(CNR)的对比度,并将它们与在传统模式下拍摄的图像的CNR进行比较。对于每种模式,在100 kVP(2.8mas),120 kVp(1.9mas)和140kvp(1.42mas)的三种曝光条件下进行三张图像。 1.61cm厚的对比细节幻像用作成像物体。对于相位对比度,对图像检测器距离(SID)的源极为182.88cm,源对象(SOD)距离为73.15厘米。 SOD与传统成像模式中的SID相同。计算的射线照相(Cr)板用作检测器,并将输出CR图像与入射X射线暴露相对地转换成线性形式。为了计算CNR,使用图像处理软件来确定兴趣区域(ROI)和ROI周围附近背景中的像素的平均像素值和标准偏差。在本研究中研究的任何给定的暴露条件下,与相应的传统模式图像相比,相位对比图像的CNR值更好。在CNR方面的卓越图像质量由相移导致的对比度贡献,以及由于物体与检测器之间的气隙导致的减少的散射。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号