【24h】

Hexagonal Surface Coil for 1.5T MRI Scanner

机译:1.5T MRI扫描仪的六边形表面线圈

获取原文

摘要

Radio-frequency coil is the core component used in the detection of nuclear magnetic resonance signal. Usually the radio frequency coil is designed for operation on a specific magnetic field strength based magnetic resonance imaging system. This paper describe the method of implementing a hexagonal surface type receive only radio frequency coil which has ability to detect the nuclear magnetic resonance signal of ~1H nuclei at 63.87 MHz resonance frequency for 1.5T magnetic resonance imaging scanner. The RF coil in MRI is used for excitation of the magnetization and to receive the signal from the excited spins. A surface coil is usually a loop of conducting material designed specifically for localized body regions. These coils have high SNR and uniform sensitivity. The Return loss of fabricated surface coil is 24.521 dB and bandwidth is 0.188 MHz.
机译:射频线圈是用于检测核磁共振信号的芯部件。通常,射频线圈设计用于在特定磁场强度的磁共振成像系统上操作。本文描述了实施六边形表面类型仅接收射频线圈的方法,该射频线圈具有在63.87MHz的谐振频率下检测〜1H核的核磁共振信号的能力,用于1.5T磁共振成像扫描仪。 MRI中的RF线圈用于激发磁化,并从激发旋转接收信号。表面线圈通常是用于专门用于局部体积的导电材料的环。这些线圈具有高SNR和均匀的灵敏度。制造的表面线圈的回流损耗为24.521dB和带宽为0.188 MHz。