The group at University of Illinois has developed the first generation MRI compatible SPECT (MRC-SPECT-I) using 40 highly pixelated CdTe detectors. In the paper, we examined strong magnetic distortion over detector charge collection, system response, and the corresponding correction model. The system capability of working in simultaneous SPECT/MRI imaging was also demonstrated using point source and resolution phantom studies. Our results showed MRC-SPECT-I could obtain sub-500μm resolution.
机译:伊利诺伊大学集团使用40个高度像素化CDTE探测器开发了第一代MRI兼容SPECT(MRC-SPECT-I)。在本文中,我们检查了对探测器充电收集,系统响应和相应校正模型的强磁失真。使用点来源和分辨率的幻影研究还证明了同时SPECT / MRI成像的系统能力。我们的结果显示MRC-SPECT-I可以获得Sub-500μm分辨率。