Defining Optimum Frequency Range for Heart Failure Detection System Considering Thickness Variations in Human Body Tissues

机译:考虑人体组织厚度变化的心力衰竭检测系统定义最佳频率范围

获取原文

摘要

One of the most apparent symptoms of heart failure that can be used for its primary detection is the accumulation of fluids in lungs (pulmonary edema). However, the competency of the non-invasive microwave detection system on varying tissues thickness (mainly fat and muscle) between the skin and lungs is one of the major challenges in the design of a successful microwave detection system. The penetration of microwave signals in human body tissue depends on the utilized frequency as well as the properties and thickness of the tissues. Hence, this study investigates optimum frequencies that can be used in monitoring the accumulation of fluids in the lungs for different thicknesses of tissues layers in the torso area. The investigation uses an equivalent circuit model of the tissues in the torso area with properties that are emulated using the fourth-order Cole-Cole model. To achieve reasonable resolution, the current study focuses on the frequency band from 0.5 GHz to 1.5 GHz. A conclusion is made from all the utilized combinations in the study that the optimum band for the investigated cases extends from 0.5 GHz to 0.9 GHz.
机译:可用于其主要检测的心力衰竭最明显的症状之一是肺部液体(肺水肿)的积累。然而,非侵入性微波检测系统对不同组织厚度(主要脂肪和肌肉)的无侵入式微波检测系统是在成功的微波检测系统设计中的主要挑战之一。微波信号在人体组织中的渗透取决于使用频率以及组织的性质和厚度。因此,本研究调查了可用于监测躯干区域中不同厚度的肺部流体积聚的最佳频率。该研究使用躯干区域中的组织的等效电路模型,其具有使用四阶COLE-COLE模型仿真的性质。为了实现合理的解决方案,目前的研究重点介绍0.5 GHz至1.5 GHz的频段。从研究中的所有使用组合进行了结论,即所研究病例的最佳带延伸到0.5GHz至0.9GHz。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号