DOSIMETRIC CHARACTERIZATION OF AN ACCESSORY MOUNTED MINI-BEAM COLLIMATOR ACROSS MULTIPLE CLINICALLY BEAM-MATCHED MEDICAL LINEAR ACCELERATORS

机译:多临床匹配医用线性加速器辅助安装迷你梁准直器的装载式微束准直

获取原文

摘要

A mini-beam collimator has been dosimetrically characterized on three clinically beam matched medical linear accelerators. Beam quality, peak-to-valley dose ratio, collimator factor, and relative output were obtained for mini-beam collimated 6 MV fields of various sizes on three Varian iX medical linear accelerators. Though the beam quality remained constant, discrepancies in peak-to-valley dose ratio, collimator factor, and relative output were observed across accelerators. Monte Carlo simulations of the mini-beam collimated fields revealed that these differences were related to the linac electron beam width.
机译:迷你束准直器在三个临床束匹配的医用线性促进剂上进行了多频性地表征。光束质量,峰谷剂量比,准直器因子和相对输出是在三个Varian IX医疗线性加速器上进行微小梁准直的6 MV领域的迷你束准直的6mV领域。虽然光束质量保持恒定,但在加速器上观察到峰谷剂量比,准直器因子和相对输出的差异。迷你梁准直场的蒙特卡罗模拟显示这些差异与LinaC电子束宽度有关。