【24h】

Extended 'Nose-to-Nose' Calibration Procedure

机译:延长“鼻向鼻子”校准程序

获取原文

摘要

When the bandwidths of oscilloscopes are similar, three different broadband sampling oscilloscopes can be calibrated by "Nose-to-Nose" calibration procedure. So far the report that "Nose-to-Nose" calibration procedure is applicable or not when bandwidth differences are relatively large has not been found. In this paper, simulations were firstly performed on "Nose-to-Nose" calibration procedure when bandwidth of oscilloscopes differ by one time, then broadband sampling oscilloscopes which bandwidth are 50GHz and 26.5GHz were studied by "Nose-to-Nose" calibration procedure. These results all indicate that "Nose-to-Nose" calibration procedure is also applicable when bandwidth differences are relatively large.
机译:当示波器的带宽相似时,可以通过“鼻到鼻子”校准程序校准三种不同的宽带采样示波器。到目前为止,“鼻向鼻子”校准程序的报告是适用的,或者在尚未找到带宽差异时,尚未找到。在本文中,当示波器的带宽一直不同,首先对“鼻到鼻子”校准程序进行仿真,然后通过“鼻到鼻子”校准程序研究了带宽的宽带采样示波器和26.5GHz的宽带采样示波器。这些结果都表明,当带宽差异相对较大时,“鼻向鼻子”校准程序也适用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号