AN ANALYSIS OF HOME AND HOSPITAL MEDICAL DEVICE INCIDENTS IN THE MAUDE DATABASE

机译:摩尔数据库中的家庭医院医疗器械事件分析

获取原文
获取外文期刊封面目录资料

摘要

A review of the U.S. Food and Drug Administration's Manufacturer and User Facility Device Experience Database (MAUDE) was performed to identify medical device incident trends relating to the type of device involved, incident location, injury severity, the nature and/or cause of the event, and human factors issues involved. A total of 317,931 medical device incident reports submitted by manufacturers between the years 1997 and 2002 were reviewed. The analysis revealed that 203,021 (64%) of these reports did not contain useful event location information. Of the remaining reports, 102,303 (89%) incidents occurred in hospitals, 6,153 (5%) incidents occurred in outpatient treatment facilities, 2,511 (2%) incidents occurred in ambulatory surgical facilities, and 3,157 (3%) incidents occurred at home. The majority of the incidents that occurred at home involved blood glucose meters (42%), defibrillators (10%), ventilators (8%), and wheelchairs (5%). Home incidents resulted in 108 fatalities and 776 injuries. The majority of the fatalities were associated with defibrillators (32%), ventilators (19%), oximeters (8%), and hemodialysis devices (6%). Seven of the 21 fatal home ventilator-related incidents were attributed to low, delayed, or non-sounding alarms. Hospital incidents comprised 102,303 records, with 2,046 incidents resulting in fatalities. A large share of the fatal hospital incidents were related to defibrillators (11%), catheters (11%), heart valves (10%), pacemakers (5%), and ventilators (4%). Six fatal hospital ventilator-related incidents were attributed in part to delayed, silenced, or non-sounding alarms.
机译:对美国食品和药物管理局的制造商和用户设施设备体验数据库(Maude)的审查是为了识别与所涉及的设备类型,事件,伤害严重程度,事件的性质和/或原因有关的医疗器械事件趋势和涉及的人为因素问题。综述了制造商之间的317,931名医疗器械事件报告,于1997年和2002年之间提交。分析显示,这些报告的203,021(64%)不包含有用的事件位置信息。在剩余的报告中,在医院发生的102,303(89%)事件发生,门诊治疗设施的6,153名(5%)发生,20,511(2%)发生在外科手术设施中发生,3,157名(3%)发生在家里发生。大多数在家中发生的事件涉及血糖米(42%),除颤器(10%),呼吸机(8%)和轮椅(5%)。家庭事件导致108人死亡和776人受伤。大多数死亡事故与除颤器(32%),呼吸机(19%),血氧仪(8%)和血液透析装置(6%)有关。 21个致命的家庭呼吸机相关事件中的七个归因于低,延迟或非探测警报。医院事件包括102,303条记录,其中2,046名事件导致死亡。大量致命的医院事件与除颤器(11%),导管(11%),心脏瓣膜(10%),起搏器(5%)和呼吸机(4%)有关。六种致命的医院呼吸机相关事件部分归因于延迟,沉默或非探测警报。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号