首页> 外文会议>International Conference on Flow Measurement >The Low Liquid Flow Calibration Facility
【24h】

The Low Liquid Flow Calibration Facility

机译:低液体流动校准设施

获取原文

摘要

The low liquid flow calibration facility is developed together by Beijing Aerospace Propulsion Institute, National Institute of Metrology and Dandong Best Automatic Engineering & Meter Ltd. The facility is designed using static-gravimetric method to calibrate mass flow meters, turbine flow meters and vortex shedding flow meters, from 4 mm to 25 mm in diameter, at flow rates ranging from 8.2 kg/h to 10890 kg/h. It is designed as a integrated movable machinery, to construct a closed system to certificate flow meters. Besides, it can be combined with special test-bed systems to certificate flow meters on-line. The facility can automatically process flow adjusting, diverter switching, water tank bleeding, data acquiring, recording and printing, etc. It incorporates a diverter valve design, which greatly reduces the uncertainty associated with the flow diversion into the collection tank. This paper details design and construction novelties of the system and outlines the uncertainty of the calibration facility.
机译:低液体流动校准设施是由北京航空推进机构,国家计量研究所和丹东最佳自动工程和仪表有限公司开发的。该设施采用静态重量法设计来校准质量流量计,涡轮机流量计和涡流流动流动米,直径为4毫米至25毫米,流速范围为8.2千克/小时至10890千克/小时。它设计为集成的可移动机械,将封闭系统构建到证书流量计。此外,它可以与特殊的测试床系统结合在线上的证书流量计。该设施可以自动处理流量调节,转向器切换,水箱出血,数据获取,记录和印刷等。它采用了分流阀设计,这大大降低了与流量转移到收集罐中的不确定性。本文详细说明了系统的设计和施工Noveltize,并概述了校准设施的不确定性。