Retrieval Analysis of a Cementless Modular Total Hip Arthroplasty Prosthesis

机译:粘土模块化总髋关节置换术假体检索分析

获取原文
获取外文期刊封面目录资料

摘要

This paper reports the analysis of a modular S-ROM hip implant, which was retrieved at revision, secondary to aseptic loosening approximately after 24 months of implantation. Light microscopy and scanning electron microscopy confirmed the presence of pitting, fretting corrosion, plastic deformation, and stress induced corrosion cracking. Energy dispersive X-ray microanalysis of the stem surface revealed the release of metal ions secondary to environmental interaction. High resolution inductively coupled plasma mass spectroscopy performed on the blood serum sample confirmed the elevated titanium ion levels (4.66 ppb). Finite element analysis of the implant showed that the micromotion and stress levels were the maximum at the proximal lateral region of the taper junction, in congruence with the observation. This case shows that taper performance is influenced by implant's metallurgy and geometry. Patient weight, activity level, and effect of the local microenvironment also play a significant role.
机译:本文报道了对修正的模块化S-ROM髋关节植入物的分析,其在植入24个月后大约在植入后的无菌松动。光学显微镜和扫描电子显微镜确认存在点蚀,微动腐蚀,塑性变形和应力诱导的腐蚀开裂。茎表面的能量分散X射线微分析显示,释放二次的环境相互作用。在血液血清样品上进行的高分辨率感应耦合等离子体质谱证实升高的钛离子水平(4.66ppb)。植入物的有限元分析表明,微调和应力水平在锥形交界处的近侧侧面的最大值,同时观察。这种情况表明,锥形性能受植入物冶金和几何的影响。患者体重,活动水平和局部微环境的影响也发挥着重要作用。