【24h】

NEW METHOD CALIBRATION FOR CALIBRATION OF A BODY BOX

机译:用于校准身体箱的新方法校准

获取原文

摘要

A rational method for the calibration or validation of Body Boxes (BB), which are used for garment cleanliness measurement in accordance to IEST-RP-CC003.3, is presented here. Without such calibration or validation the Body Boxes may be space biased and may not even be valid for comparative evaluation of garments. The method presented here involves point source aerosol generation within a tube, dilution and comparison of calculated dilution concentrations, based on tube concentration measurements, to actual measured BB sampler values. The ratio of these two values then results in the BB calibration factor for that point or zone within the BB. By repeating the process an average calibration value and variation or range is determined. Results of two such validations of actual BBs are presented in the paper.
机译:本文介绍了根据IEST-RP-CC003.3校准或验证身体箱(BB)的校准或验证的合理方法,其用于衣服清洁度测量。在没有这种校准或验证的情况下,身体箱可以是空间偏置,并且甚至可能甚至不适用于服装的比较评估。此处呈现的方法涉及基于管浓度测量的管,稀释和比较计算的稀释浓度的点源气溶胶产生,以实际测量的BB采样器值。然后,这两个值的比率导致BB内的BB校准因子或BB内的区域。通过重复过程,确定平均校准值和变化或范围。在纸上提出了两个这样的实际BBS验证的结果。