CALIBRATION ACCURACY OF THE LONG STEM SPRT IN THE TEMPERATURE RANGE -195°C TO 0°C USING COMPARISON METHOD AND BOILING POINT OF NITROGEN

机译:使用比较法和氮气沸点校准温度范围为-195℃至0°C的长杆SPRIC校准精度

获取原文

摘要

The ITS-90 stipulates the triple point of Argon as defining fixed point in the range below 273 K. In theory this fixed point is very simple to establish, but to realize it in the laboratory implies the equipment necessary, consisting of one argon sealed cell and a cryostat. This equipment is expensive and it requires a long period of time to establish the triple point temperature. An alternative used by many calibration laboratories is the boiling point of nitrogen. An investigation on the use of the boiling point of nitrogen and the accuracy of the calibration was performed with the help of a commercially available equipment and instrumentation. A serial of Standard Platinum Resistance Thermometers were calibrated by fixed point method and the interpolation function of ITS-90. The same thermometers were then calibrated in the same temperature range by using the boiling point of nitrogen and liquid calibration bath. Different methods were adopted and the results were compared, including an investigation of the measurement uncertainties.
机译:其-90规定了氩的三重点,如在273 k的范围内定义固定点。理论上,这种定点非常简单,但在实验室中实现它意味着由一个氩密密封的设备和一个低温恒温器。该设备昂贵,需要很长一段时间来建立三重点温度。许多校准实验室使用的替代方案是氮的沸点。在市售的设备和仪器的帮助下,执行关于氮气沸点的使用和校准精度的研究。通过固定点法和ITS-90的插值函数校准了一系列标准铂电阻温度计。然后通过使用氮气和液体校准浴的沸点在相同温度范围内校准相同的温度计。采用不同的方法,比较结果,包括对测量不确定性的调查。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号