首页> 外文会议>International SYmposium on Temperature and Thermal Measurements in Industry and Science >EFFECT OF PURITY OF THE PLATINUM WIRE OF HTPRTS ON THEIR CHARACTERISTICS
【24h】

EFFECT OF PURITY OF THE PLATINUM WIRE OF HTPRTS ON THEIR CHARACTERISTICS

机译:HTPRTS铂金属焊丝纯度对其特征的影响

获取原文

摘要

High temperature platinum resistance thermometers (HTPRTs) were introduced in the International Temperature Scale of 1990, replacing thermocouples in the range 660.323°C - 961.78°C. That resulted in increasing the reproducibility of the scale by more than an order of magnitude. In the definition of ITS-90 there is a requirement on the quality of the platinum used in the platinum resistance thermometers in all the temperature sub-ranges above 0°C. The resistance ratio, W(Ga), at the melting point of gallium must not be less than 1.11807. Experience in using HTPRTs showed that different long-term stability of the resistance might be obtained at different temperatures. The reasons for less stability at high temperatures are possibly due to changes in crystal structure of the platinum and mobility of impurities and defects. In this paper the work, carried out at VNIIM on the investigation of HTSPRTs with different values of W(Ga) is analyzed and summarized to discuss the influence of the purity of the platinum wire on the stability of the HTSPRTs. A suggestion is made to tighten the requirement on W(Ga) for platinum resistance thermometers used at high temperature as standard interpolation instruments of the ITS-90.
机译:高温铂电阻温度计(HTPRTS)在1990年的国际温度范围内引入,更换了660.323°C-961.78°C范围内的热电偶。这导致规模的再现性增加了超过一个数量级。在其-90的定义中,在0℃高于0℃的所有温度子范围内,铂电阻温度计中使用的铂的质量有所要求。镓在熔点的电阻比W(Ga)不得小于1.11807。使用HTPRTS的经验表明,在不同的温度下可以获得不同的长期稳定性。在高温下稳定性较低的原因可能是由于铂的晶体结构的变化和杂质和缺陷的迁移率。在本文中,分析了在VNIIM上进行的工作,并概述了不同W(GA)的HTSPTTS的研究,并概述讨论铂金线纯度对HTSPRTS稳定性的影响。提出建议,拧紧高温下使用的铂电阻温度计的W(Ga)的要求,作为其-90的标准插值仪器。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号