Following an earlier characterization of the PSB frost point generator [1], the generator was further characterized by using a high precision chilled mirror dew point meter. The characterization includes verification of the generator's saturation efficiency at different combinations of saturation pressure and temperature and the influence of the flow rate used. Comparisons between the generator and the dew point meter were also carried out in the full operating range of the generator.
展开▼