Dosimetric characterization of a new /sup 125/Iodine brachytherapy source, model I125-SL

机译:新型/ SUP 125 /碘近距离放射治疗源,模型I125-SL的剂量表征

获取原文

摘要

Low-energy gamma emitting isotopes encapsulated for permanent implant are routinely applied in prostate cancer brachytherapy. Before clinical use, a new source design requires full dosimetric analysis and calibration standardization. The results of one such experimental measurement and analysis are here reported for a new design of /sup 125/Iodine source, model I125-SL (Mills Biopharmaceutical Inc.). Dose measurements were made using standard methods employing thermoluminescent dosimeters in a water equivalent plastic phantom, in accord with the AAMP Task Group 43 recommendation of liquid water reference material. Precision machined bores in the phantom located dosimeters and source(s) in a reproducible fixed geometry providing for transverse-axis and angular dose profiles over a range of distances from 0.17 to 10 cm. The data were analyzed in terms of parameters recommended by AAPM TG43. The dose-rate constant, /spl Lambda/, was evaluated by two methods, the first with reference to a /sup 60/Cobalt standard, accounting for response variation with isotope energy spectrum. Second, the dose-rate constant was determined with reference to phantom measurements using calibrated model 6702 and 6711 sources. The radial dose function, g(r), the anisotropy function, F(r,/spl theta/), the anisotropy factor, /spl phi//sub an/(r), and the point-source approximation anisotropy constant, /spl phi/~/sub an/, were derived from one- and two-dimensional dose distribution data measured in the phantom, accounting for finite dosimeter volume and with attention to inter-chip effects. The results are compared to TG43 data for existing /sup 125/I sources. The new source is comparable to the model 6711 source design.
机译:低能量的γ发射封装永久植入物的同位素在前列腺癌近距离放射治疗使用常规应用于。临床使用之前,新中源的设计需要充分的剂量学分析和校准标准化。一个这样的实验测量和分析的结果在这里报道/ SUP 125 /碘源,模型I125-SL(米尔斯生物制药公司)的一种新的设计。剂量测量是使用采用热释剂量计在水等效塑料体模的标准方法制成,在与液体水的参考材料的AAMP任务组43建议一致。精密加工在体模位于剂量计并以可重复的固定几何形状从0.17到10厘米的范围内的距离的提供横向轴和角剂量分布源(一个或多个)孔。由AAPM TG43推荐的参数方面的数据进行了分析。剂量率常数,/ SPL LAMBDA /,通过两种方法中,第一参照/ SUP 60 /钴标准,占用同位素能谱响应变化进行评价。其次,剂量率常数是参考使用校准的模型6702和6711个源幻影测量来确定。所述径向剂量函数g(R),所述各向异性函数F(R,/ SPL THETA /)时,各向异性因子/ SPL披//子的/(R),以及所述点源近似各向异性常数,/ SPL PHI /〜/分的/中,从在体模测量一维和两维剂量分布数据,占有限剂量计体积,并用注意芯片间的效果的。结果相比,现有的/ SUP 125 / I源TG43数据。新源媲美的型号6711源设计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号