Two methods, one approximate and one theoretically (mathematically) exact, for calculation of parameters of measured low frequency sinusoidal signal are thoroughly studied and compared. Both methods perform sampling of measured signal by HP 3458A Digital Multimeter in DC mode in order to achieve the best possible accuracy. Comparison was made firstly on calculated, ideal sinusoid (computer simulation) what has shown us theoretical limits of accuracy due to calculating inaccuracy and approximations. Then a series of measurements in real, laboratory conditions was made to confirm or refute theoretical foresights. For both methods, advantages and disadvantages are noticed and main influences on the final result uncertainty are explained.
机译:彻底研究并比较了两种方法,用于计算测量的低频正弦信号的参数的一个近似和理论上(数学上)精确的。两种方法在DC模式下通过HP 3458A数字万用表进行测量信号的采样,以实现最佳的精度。首先进行了比较,理想的正弦曲线(计算机仿真)由于计算不准确和近似而引起了对我们的理论极限的理论限制。然后是真实的一系列测量,进行了实验室条件,以确认或反驳理论远见。对于这两种方法,还注意到了优点和缺点,并解释了对最终结果不确定性的主要影响。