A device for measuring soft tissue pressure: application to human intervertebral disc and bovine tendon

机译:一种用于测量软组织压力的装置:对人椎间盘和牛肌腱的应用

获取原文

摘要

A needle mounted strain gauge pressure sensor has been developed that has proven to be both accurate and reliable. It is a resistive strain gauge mounted on a pressure diaphragm which is exposed to pressure through a silicone window. The transducer is used in a Wheatstone bridge configuration and incorporates a dummy resister for temperature compensation. The transducer cable also contains an excitation voltage limiting circuit for protection of the sensor. The authors used a Measurements Group model 2100 amplifier to provide an excitation voltage of 3 V DC and a gain of 300. Although the sensor's maximum deviation from linearity is over 5%, its curve can be fitted by a quadratic to within 0.4% error. Its use for intradiscal pressure measurements appears to be very satisfactory. Measurement of tendon and ligament tensile force was also good.
机译:已经开发了针头安装的应变压力表压力传感器,其已被证明是准确可靠的。它是安装在压力膜片上的电阻应变计,其通过硅树脂窗口暴露于压力。换能器用于惠斯通桥配置,并结合一个用于温度补偿的虚设电阻器。换能器电缆还包含用于保护传感器的激励电压限制电路。作者使用了测量组模型2100放大器,提供了3 V DC的激励电压,并且增益为300.虽然传感器的线性最大偏差超过5%,但其曲线可以在0.4%误差范围内拟合。它用于盘内压力测量的用途似乎非常令人满意。肌腱和韧带拉伸力的测量也很好。