首页> 外文会议>Specialist Meeting on Microwave Radiometry and Remote Sensing for the Environment >Analysis of Tipping-Curve Acceptance Criteria for Ground-Based Radiometer Calibration
【24h】

Analysis of Tipping-Curve Acceptance Criteria for Ground-Based Radiometer Calibration

机译:地面辐射计校准的划线曲线验收标准分析

获取原文
获取外文期刊封面目录资料

摘要

The tipping-curve method is widely used to provide the cold calibration point for calibration of ground-based radiometers. Statistical analysis of the opacity air mass pairs have been used as measures of the tip curve quality, but the choice of acceptance criteria has been subjective. An analysis of the tipping curve acceptance criteria was accomplished by simulation of a total power radiometer under non stratified conditions due to perturbations of water vapor content. Relationships between the root mean square error (RMSE) in zenith brightness temperature and regression coefficient were examined. While the RMSE decreased with increasing regression coefficient, a stringent acceptance criteria did not ensure accuracy likely due to the presence of net horizontal gradients. On the other hand, for radiometers with multiple channels the standard deviation of brightness temperature measurements of a reference black body, also known as spectral consistency, was found to be a robust indicator of calibration accuracy.
机译:倾翻曲线方法广泛用于提供用于校准地基辐射仪的冷校准点。不透明度空气质量对的统计分析已被用作尖端曲线质量的衡量标准,但接受标准的选择是主观的。通过对水蒸气含量的扰动模拟非分层条件下的总功率辐射计来实现倾翻曲线验收标准的分析。检查了Zenith亮度温度和回归系数的根均方误差(RMSE)之间的关系。虽然RMSE随着回归系数的增加而降低,但严格的验收标准没有确保由于净水平梯度的存在而可能的可能性。另一方面,对于具有多个通道的辐射仪,发现参考黑色主体的标准偏差,也称为光谱一致性,是校准精度的强大指标。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号