Non-contact measurements of diaphragm electromyogram, electrocardiogram and respiratory variations with sheet-type fabric electrodes for neonatal monitoring

机译:片状织物电极用于新生儿监护的隔膜肌电图,心电图和呼吸变化的非接触式测量

获取原文

摘要

Herein, a sensor system was developed for non-contact electrocardiogram (NcECG), non-contact diaphragm electromyogram (NcEMGdiaph), and NcEMGdiaph-derived respiratory variation (NcRVEMG) measurements that use sheet-type fabric electrodes placed under a subject in a supine position. We developed prototype system was evaluated with both adult and infant subjects. In a basic evaluation test on the adult subject, we confirmed that the reference respiration rates and the NcRVEMG signals were highly correlated (r = 0.96). Tests with three infant subjects confirmed that the prototype system can simultaneously measure the NcECG, NcEMGdiaph, and NcRVEMG signals. In these tests, the proposed system exhibited improved sensitivity by 19.7% and accuracy by 10.5% These results suggested that the developed prototype system can measure the NcECG, NcEMGdiaph, and NcRVEMG signals in infants during sleep or at near-complete rest.
机译:在此,针对非接触式心电图(NcECG),非接触式隔膜肌电图(NcEMG diaph )和NcEMG diaph 派生的呼吸变化(NcRV < sub> EMG )测量,该测量使用了将薄片型织物电极以仰卧位放置在受试者下方的方法。我们开发的原型系统已针对成人和婴儿受试者进行了评估。在对成人受试者的基本评估测试中,我们确认参考呼吸频率与NcRVEMG信号高度相关(r = 0.96)。对三名婴儿受试者的测试证实,原型系统可以同时测量NcECG,NcEMG diaph 和NcRV EMG 信号。在这些测试中,拟议的系统显示出提高了19.7%的灵敏度和10.5%的准确性。这些结果表明,开发的原型系统可以测量婴儿在睡眠或接近完全休息时的NcECG,NcEMGdiaph和NcRVEMG信号。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号