The appearances of Motion Artifacts (MAs) in the photoplethysmographic (PPG) signals is one of the major obstacles to improve the accuracy and the stability of the signals analysis. In this paper, we present an effective and adaptive method based on DC Remover method and Recursive Least Squares (RLS) adaptive filter for reducing MAs disturbances from PPG signals. The results achieved by the presented methodology show a high correlation coefficient between Electrocardiography derived heart rate and PPG-derived heart rate (R=0.8054), moreover, the accuracy of heart rate monitoring was improved. The proposed system was implemented in hardware design for wearable and home-care applications.
机译:在光电容积描记器(PPG)信号中出现运动伪像(MAs)是提高信号分析的准确性和稳定性的主要障碍之一。在本文中,我们提出了一种基于DC去除器方法和递归最小二乘(RLS)自适应滤波器的有效且自适应的方法,用于减少来自PPG信号的MAs干扰。通过本方法获得的结果表明,心电图得出的心率与PPG衍生的心率之间具有较高的相关系数(R = 0.8054),而且,心率监测的准确性得到了提高。拟议的系统已在可穿戴和家庭护理应用的硬件设计中实现。