We describe a visible-light multi-spectral system for vascular oximetry studies that can be implemented in low-and middle-income countries, using a low-cost electronics and optical elements, for instance a Raspberry Pi, aPi camera under a resolution of 5-megapixel, and four light sources at 480nm, 532nm, 593nm and 610nm. Itis designed to quantify the vascular oxygen saturation of the rat dorsal spinal cord, Python custom applicationsynchronizes all elements to execute the imaging process in one system. The device is suitable offers a replacementfor high cost bulky systems in drug discovery, tracking disease progression and understanding of progressive anddegenerative diseases.
机译:我们描述了一种用于血管血氧饱和度研究的可见光多光谱系统,该系统可以在低 和中等收入国家/地区,它们使用低成本的电子和光学元件,例如Raspberry Pi, Pi相机的分辨率为5百万像素,并具有480nm,532nm,593nm和610nm的四个光源。它 专为量化大鼠背脊髓的血管氧饱和度而设计,Python自定义应用程序 同步所有元素以在一个系统中执行成像过程。该设备适合提供替换 用于昂贵的笨重系统中的药物发现,跟踪疾病进展以及对渐进性疾病的了解 退行性疾病。