首页> 外文会议>International Instrumentation and Measurement Technology Conference >Tank-tests of a Prototype Electromagnetic Groundwater Flowmeter
【24h】

Tank-tests of a Prototype Electromagnetic Groundwater Flowmeter

机译:原型电磁地下水流量计的水箱测试

获取原文

摘要

Nutrients leaching through groundwater are one contributor to waterway contamination. Guiding land use decisions to reduce the impact on water resources requires accurate knowledge of the flow of groundwater. Currently, measurements from monitored wells are used to tune models of this flow. However, there are few wells available and so the data is sparse. An electromagnetic flowmeter has been proposed to simplify the measurement of groundwater flow and provide data to generate more robust models. Although based on a sensing technique that has been widely used for measurement of flow in pipes, the slow movement of groundwater results in small signals subject to significant interference. This paper presents some initial laboratory results from a prototype electromagnetic flowmeter showing that there is a tradeoff between noise and self-interference reduction.
机译:通过地下水浸出的营养素是水道污染的贡献者。指导土地利用决定减少对水资源的影响需要准确地了解地下水的流动。目前,受监控井的测量用于调整该流程的模型。但是,很少有井可用,因此数据稀疏。已经提出了一种电磁流量计来简化地下水流量的测量,并提供数据以产生更强大的模型。尽管基于广泛用于测量管道的流量的传感技术,但地下水的缓慢运动导致小信号受到显着干扰。本文提出了来自原型电磁流量计的一些初始实验室结果,表明噪声和自干扰之间存在权衡。