首页> 外文会议>Engineering in Medicine and Biology Society, 1989. Images of the Twenty-First Century., Proceedings of the Annual International Conference of the IEEE Engineering in >An instrument for measuring the absolute output acoustic intensity,effective radiating area and beam non-uniformity ratio of medicalultrasound devices
【24h】

An instrument for measuring the absolute output acoustic intensity,effective radiating area and beam non-uniformity ratio of medicalultrasound devices

机译:一种用于测量绝对输出声强的仪器,医用有效辐射面积与光束不均匀率超声波设备

获取原文

摘要

A computer-controlled acoustic intensity measuring system (AIMS)for measuring the acoustic output of commercially manufactureddiagnostic and therapeutic ultrasound devices is described. The hardwareand software performance criteria and design tradeoffs resulting fromexperience with biomedical sound fields are also presented. The systemallows three-orthogonal-axis (X, Y, and Z)position translation of up to 10 in by 10 in by 20 in with a 6.5-μmminimum step size, and is capable of waveform acquisition with a 50-MHzanalog bandwidth and a maximum equivalent time sample rate of 5 GHz. Theinstrument can perform planar scans wherein the scan plane contains theacoustic beam axis, and two polar axes of position rotation can be addedto the main positioner assembly at a later date
机译:计算机控制的声强测量系统(AIMS) 用于测量商业制造的声音输出 描述了诊断和治疗超声设备。硬体 以及软件性能标准和设计折衷所产生的 还介绍了生物医学声场的经验。系统 允许三个正交轴( X Y Z ) 6.5μm的位置平移可达10 in x 10 in x 20 in 最小步长,并能够以50MHz的频率采集波形 模拟带宽和5 GHz的最大等效时间采样率。这 仪器可以执行平面扫描,其中扫描平面包含 声束轴和位置旋转的两个极轴可以添加 稍后再发送到主定位器组件