首页> 外文会议>Medical Imaging Conference >Preliminary imaging evaluation of a compact tomosynthesis system for potential point-of-care extremity imaging
【24h】

Preliminary imaging evaluation of a compact tomosynthesis system for potential point-of-care extremity imaging

机译:紧凑型断层合成系统对潜在即时医疗点肢端成像的初步成像评估

获取原文

摘要

Orthopedic tomosynthesis is emerging as an attractive alternative to digital radiography (DR), with increased sensitivity for some clinical tasks, including fracture diagnosis and staging and follow-up of arthritis. Commercially available digital tomosynthesis (DTS) systems are complex, room-sized devices. A compact tomosynthesis system for extremity imaging (TomoE) was previously demonstrated using carbon nanotube (CNT) x-ray source array technology. The purpose of this study was to evaluate the prototype device in preparation for an Institutional Review Board (IRB)-approved patient imaging study and evaluate initial patient images.A tabletop device was constructed using a short CNT x-ray source array, operated in three positions, and a flat panel digital detector. Twenty-one x-ray projection images were acquired at incident angles from -20 to +20 degrees in various clinical orientations, with entrance doses matched to commercial in-room DTS scanners. Theaprojection images were reconstructed with an iterative reconstruction technique in 1mm slices. Cadaveric specimen and initial participant images were reviewed by radiologists for feature conspicuity and diagnostic accuracy.TomoE image quality was found to be superior to DR, with reconstruction slices exhibiting visual conspicuity of trabecular bone, delineation of joint space, bone erosions, fractures, and clear depiction of normal anatomical features. The scan time was fifteen seconds with mechanical translation. Skin entrance dose was verified to be 0.2mGy.TomoE device image quality has been evaluated in cadaveric specimens and dose was calibrated for a patient imaging study. Initial patient images depict a high level of anatomical detail an increase in diagnostic value compared to DR.
机译:骨科断层合成正在成为数字放射成像(DR)的一种有吸引力的替代方法,它对某些临床任务(包括骨折诊断,关节炎分期和随访)的敏感性提高。市售的数字断层合成(DTS)系统是复杂的,房间大小的设备。先前已使用碳纳米管(CNT)X射线源阵列技术演示了用于四肢成像(TomoE)的紧凑型断层合成系统。这项研究的目的是评估原型设备,以准备接受机构审查委员会(IRB)批准的患者影像学研究并评估初始患者图像。使用短CNT X射线源阵列构建台式设备,该设备应分三步操作位置和平板数字检测器。在-20到+20度的入射角下,在各种临床方向上采集了二十一张X射线投影图像,其入射剂量与商用室内DTS扫描仪相匹配。用迭代重建技术在1mm切片中重建投影图像。放射科医生审查了尸体标本和初始参与者图像的特征显着性和诊断准确性,发现TomoE图像质量优于DR,重建切片显示了小梁骨的视觉显着性,关节间隙的轮廓,骨侵蚀,骨折和清晰正常解剖特征的描述。机械平移的扫描时间为十五秒。皮肤入口剂量经验证为0.2mGy。已在尸体标本中评估了TomoE设备的图像质量,并针对患者影像学研究对剂量进行了校准。初始患者图像显示高水平的解剖细节,与DR相比,诊断价值有所提高。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号