Bilateral comparison of current transformer measuring systems of NCHVM and PTB

机译:NCHVM和PTB电流互感器测量系统的双边比较

获取原文

摘要

An intercomparison of current transformer calibration systems was performed between National Center For High Voltage Measurement (NCHVM) and Physikalisch-Technische Bundesanstalt (PTB) in order to confirm their measurement capabilities. A current transformer was used as travelling standard with transformer ratios of 5A/5A, 10A/5A and 100A/5A at a frequency of 50 Hz and at ambient temperatures of $23^{circ}mathrm{C}pm 3^{circ}mathrm{C}$. The measurements has been performed at test points varying from 5% to 120% of the rated current. For these three ratios, the test results differ by no more than 1.3 parts in 106 for the ratio error and $2.0 mu mathrm{rad}$ for the phase displacement.
机译:为了确定其测量能力,在国家高压测量中心(NCHVM)和物理技术联合会(PTB)之间进行了电流互感器校准系统的比较。电流互感器用作行进标准,在50 Hz的频率和环境温度为50 Hz时的互感器比率为5A / 5A,10A / 5A和100​​A / 5A。 $ 23 ^ {\ circ} \ mathrm {C} \ pm 3 ^ {\ circ} \ mathrm {C} $ 。测量是在额定电流的5%至120%的测试点上进行的。对于这三个比率,测试结果相差不超过1.3十分之几 6 比率误差和 $ 2.0 \ \ mu \ mathrm { rad} $ 用于相移。