摘要

An incremental buildup calibration method of inductive voltage divider was introduced, which comprised two steps, the zero voltage measurement procedure and the section voltage measurement procedure. The reference transformer is connected in series with an auxiliary IVD to compare voltage with the output of IVD to be calibrated. The influence of leakage current of both the reference transformer and the detection transformer can be removed by this method.
机译:介绍了一种增量式电感分压器的标定方法,该方法包括两个步骤:零电压测量程序和截面电压测量程序。参考变压器与辅助IVD串联连接,以将电压与要校准的IVD输出进行比较。通过这种方法可以消除参考变压器和检测变压器的漏电流的影响。