Objective assessment of the effect of Cardiac Resynchronization Therapy (CRT) still represents a very difficult clinical task. The ultra-high-frequency electrocardiographic (ECG) dyssynchrony (e-DYS) represents a simple ECG derived method that provides information about the ventricular electrical depolarization and thus could become helpful in CRT. The 12-lead ECG (sampling 5kHz/24bits) and echocardiography were obtained in 42 consecutive CRT patients. The e-DYS and standard QRS duration (QRSd) were automatically assessed before, just after, and 6 months after the CRT during CRT Off and CRT On. In 30 CRT responders determined by echocardiography, the e-DYS parameter was significantly lower after 6 months during CRT Off compared to state prior CRT (-9±19ms, p<0.05), while there was no significant e-DYS change during CRT On within 6 months (-1±11ms, NS). A clear shortage of e-DYS was observed just after CRT On (-46±23ms, p<10e-10), and 6 months after CRT On (-46±21ms, p<10e-10) compared to state before CRT. No significant changes of the e-DYS parameter were found in 12 CRT non-responders. The QRSd didn’t show any significance in all observed changes. The e-DYS parameter seems to be a superior parameter to simply and quickly assess an immediate as well as the long-term effect in CRT.
机译:心脏再同步治疗(CRT)效果的客观评估仍然是一项非常困难的临床任务。超高频心电图(ECG)不同步(e-DYS)代表一种简单的ECG衍生方法,可提供有关心室电去极化的信息,因此可能对CRT有所帮助。在连续的42例CRT患者中获得了12导联心电图(采样频率为5kHz / 24bits)和超声心动图。在CRT关闭和CRT打开期间,在CRT之前,之后和之后6个月自动评估e-DYS和标准QRS持续时间(QRSd)。在30例通过超声心动图检查确定的CRT应答者中,CRT停用6个月后e-DYS参数显着低于先前的CRT状态(-9±19ms,p <0.05),而CRT开启期间e-DYS没有明显变化。在6个月内(-1±11ms,NS)。在CRT开启后(-46±23ms,p <10e- 10 )和CRT开启后6个月(-46±21ms,p <10e -10 )与CRT之前的状态相比。在12个CRT无响应者中未发现e-DYS参数的显着变化。 QRSd在所有观察到的变化中均未显示任何意义。 e-DYS参数似乎是一个更好的参数,可以简单,快速地评估CRT的即时效果和长期效果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号