Development of Methods and Algorithms for Determining Physiological Parameters of the Patient with the Help of the Biomodule

机译:借助生物模块确定患者生理参数的方法和算法的开发

获取原文

摘要

The purpose of this article is to develop methods and algorithms for detecting QRS complexes and cardiac episodes, primarily diagnosing tachy- and bradyarrhythmias, as well as heart pauses and atrial fibrillation. Datasets were collected by BioModule BioHarness 3 (Zephyr Technology Corporation, Annapolis, MD). The monitor was installed simultaneously with Holter monitor MYOCARD-HOLTER (NIMP ESN LLC, Sarov, Nizhny Novgorod region). Data from the PhysioNet database was used to verify the QRS detector and RR-interval analyzer. The algorithm is validated using the PhysioNet database, resulting a sensitivity of Se = 98.26% and a specificity Sp = 99.33%, and using the BioMonitor 3, resulting a sensitivity of Se = 98.95% and a specificity Sp = 93.36%. 16 records were analyzed for the presence event such as tachycardia and bradycardia, cardiac pauses and atrial fibrillation. 14 of 16 were recognized correctly by the analyzer. The analyzer gave an incorrect result in 2 cases.
机译:本文的目的是开发检测QRS复合体和心脏发作的方法和算法,主要诊断心动过速和缓慢性心律失常以及心脏停顿和房颤。数据集由BioModule BioHarness 3(Zephyr Technology Corporation,安纳波利斯,马里兰州)收集。该监视器与动态心电图监视器MYOCARD-HOLTER(下诺夫哥罗德州萨罗夫的NIMP ESN LLC)同时安装。来自PhysioNet数据库的数据用于验证QRS检测器和RR间隔分析仪。使用PhysioNet数据库验证了该算法,得出的敏感性为Se = 98.26%,特异性Sp = 99.33%,并使用BioMonitor 3进行了验证,得出的敏感性为Se = 98.95%,特异性Sp = 93.36%。分析了16个记录的存在事件,例如心动过速和心动过缓,心脏停顿和心房颤动。分析仪正确识别出16个中的14个。分析仪在2种情况下给出了错误的结果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号