首页> 外文会议>IEEE biomedical circuits and systems conference >Objective Human Gustatory Sensitivity Assessment Through a Portable Electronic Device
【24h】

Objective Human Gustatory Sensitivity Assessment Through a Portable Electronic Device

机译:通过便携式电子设备进行客观的人体味觉敏感性评估

获取原文

摘要

Objective assessment of taste perception is a challenging problem. In general, psychophysical tests based on the administration of a prototypical bitter compound, 6-n-propyltiouracil (PROP), and on a subjective scoring, are exploited. Since the PROP tasting ability is correlated to food preference and also to several clinical and health-related conditions, the development of low-cost hardware/software tools for accurate testing (e.g., in pharmacies) is an important goal. Previous works revealed the possibility to perform an electrophysiological measurement on the tongue able to provide objectiveness and repeatability in this kind of study. The technique, called Electrotastegram (ETG), relies on a custom thin-film electrode and an off-the-shelf electrophysiological recording system. In this work, we present a prototypical low-cost pocket-sized wireless device for ETG, comparing the signal quality against a top-class electrophysiological recording system already used for the same measurements. A custom Matlab interface was developed to collect and process the signal. By acquiring the signals at the same time with both devices, it was possible to compare the quality in terms of noise, amplitude and waveform. The results reveal a substantial equivalence between the two devices, thus opening to the possibility of using the proposed system in real scenarios, possibly substituting the Matlab interface with a stand-alone software.
机译:味觉的客观评估是一个具有挑战性的问题。通常,利用基于典型苦味化合物6-n-丙基-替尿嘧啶(PROP)的给药以及主观评分的心理物理测试。由于PROP品尝能力与食物偏爱以及与若干临床和健康相关的状况相关,因此开发用于精确测试(例如在药房中)的低成本硬件/软件工具是一个重要目标。先前的研究揭示了在这种研究中对舌头进行电生理测量的可能性,从而能够提供客观性和可重复性。这项技术称为电子味觉图(ETG),它依赖于定制的薄膜电极和现成的电生理记录系统。在这项工作中,我们提出了一种用于ETG的原型低成本袖珍无线设备,将信号质量与已经用于相同测量的顶级电生理记录系统进行了比较。开发了自定义的Matlab接口来收集和处理信号。通过同时使用两个设备采集信号,可以比较噪声,幅度和波形的质量。结果揭示了这两种设备之间的实质等效性,从而开辟了在实际场景中使用建议的系统的可能性,可能会用独立软件代替Matlab接口。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号