首页> 外文会议>IEEE Nuclear Science Symposium;Medical Imaging Conference >PD-HEXID1: A Low-Power, Low-Noise Pixel Readout ASIC for Pixelated-Scintillator-Based X-Ray Detectors
【24h】

PD-HEXID1: A Low-Power, Low-Noise Pixel Readout ASIC for Pixelated-Scintillator-Based X-Ray Detectors

机译:PD-HEXID1:用于基于像素化闪烁体的X射线探测器的低功耗,低噪声像素读出ASIC

获取原文

摘要

A low-power, low-noise pixel readout application-specific-integrated-circuit (ASIC) prototype with embedded in-pixel photo-diode was developed for a novel pixelated-scintillator-based X-ray detector, capable of spectral energy discrimination and suitable for low dose whole body CT imaging. The ASIC includes 8-by-16 channels to read out negative charges detected by the built-in photodiodes, at a pitch size of 250 μm to achieve good spatial and energy resolution energy of detected photons. Each channel of the ASIC provides low-noise charge amplification, high-order shaping with baseline stabilization, discrimination, extraction of amplitude including neighbor channels, multiplexing, and dissipates ~0.6 mW. The measured equivalent noise charge (ENC) is about 20 rms electrons at 500 ns peaking time. The ASIC with built-in photodiodes has demonstrated high sensitivity to light.
机译:一种具有嵌入式像素内光电二极管的低功耗,低噪声像素读出专用集成电路(ASIC)原型,用于新型基于像素化闪烁体的X射线检测器,能够检测光谱能量并适用于低剂量全身CT成像。 ASIC包括8 x 16通道,以250μm的间距读取内置光电二极管检测到的负电荷,以实现检测到的光子的良好空间和能量分辨率能量。 ASIC的每个通道都提供低噪声电荷放大,具有基线稳定度的高阶整形,辨别力,包括相邻通道的幅度提取,多路复用和耗散〜0.6 mW。在500 ns的峰值时间,测得的等效噪声电荷(ENC)为约20 rms电子。内置光电二极管的ASIC对光具有很高的灵敏度。