摘要

In modern medicine, the batteries in implantable devices are replaced with wireless power transmission (WPT) technology. The resonance-based WPT method is widely used in biomedical implantation applications as it enables the mid-field operation within the human body. The shape and misalignment of the implantable coil due to human breathing activity and body moments are the main obstacles to achieve the high power transfer efficiency. In this paper, we designed, optimized, and analyzed the spherical and circular shaped implantable coils to mitigate the misalignment issue. Results show that the spherical shaped implantable coil has superior performance under misalignment situations when compared to the circular shaped implantable coil.
机译:在现代医学中,可植入设备中的电池已被无线电源传输(WPT)技术取代。基于共振的WPT方法广泛用于生物医学植入应用,因为它可以在人体中进行中场操作。由于人的呼吸活动和身体力矩引起的可植入线圈的形状和未对准是实现高功率传输效率的主要障碍。在本文中,我们设计,优化和分析了球形和圆形植入式线圈,以减轻不对准问题。结果表明,与圆形可植入线圈相比,球形可植入线圈在未对准情况下具有优越的性能。