首页> 外文会议>IEEE International Symposium on Circuits and Systems >Live Demonstration: A 1280 by 720 by 3, 250 mW, 24 fps Hexachromatic Imager for Near-Infrared Fluorescence Image-Guided Surgery
【24h】

Live Demonstration: A 1280 by 720 by 3, 250 mW, 24 fps Hexachromatic Imager for Near-Infrared Fluorescence Image-Guided Surgery

机译:现场演示:1280 x 720 x 3、250 mW,24 fps的六色成像仪,用于近红外荧光图像引导手术

获取原文

摘要

We have designed, fabricated, and tested a hexachromatic imager for near-infrared fluorescence image-guided surgery capable of simultaneously imaging three visible and near-infrared spectral channels. The sensor is comprised of 1280 by 720 pixels yielding a 48 dB signal-to-noise-ratio and achieving a minimum indocyanine green concentration detectability of 5 nM. The imaging sensor combines the vertically stacked photodiode technology and an array of pixelated spectral interference filters in a single-chip single-snapshot architecture yielding high sensitivity and clinical translation.
机译:我们已经为近红外荧光图像引导手术设计,制造和测试了六色成像仪,该成像仪能够同时对三个可见和近红外光谱通道成像。该传感器由1280 x 720像素组成,可产生48 dB的信噪比,并具有5​​ nM的最小吲哚菁绿浓度检测能力。成像传感器在单芯片单快照架构中结合了垂直堆叠的光电二极管技术和像素化光谱干涉滤光片阵列,可实现高灵敏度和临床转换。